Informatique
EM-5 Flue Gas Analyzer |
Measuring gas Interference |
SO2 |
NO |
NO2 |
O2 |
SO2 (500ppm) |
/ |
<1ppm</p> |
no |
no |
NO (500ppm) |
no |
/ |
no |
no |
NO2 (500ppm) |
no |
<1ppm</p> |
/ |
no |
H2O (no dew) |
no |
no |
no |
no |
CO (1000ppm) |
no |
no |
no |
no |
CO2 (20%) |
no |
no |
no |
no |
O2 (21%) |
no |
no |
no |
/ |
Our EM-5 continuous flue online gas analyser hashigh reliability
Our EM-5 continuousfluegas analyzer hassmall zero drift and span drift
Our EM-5online gas analyser can do modular design
Our EM-5 continuous flue online gas analyser hasno optical moving parts, without vibration influence
Our EM-5 continuousfluegas analyzerhas strong gas chamber, low cost
Our EM-5 continuousfluegas analyzer has automatic adjustment spectrum technology, long maintenance-free period
Our EM-5online gas analyser canadopt pulse light source with 10-year service life
Technical Parameter Of EM-5 Flue Gas Analyzer
Measurement Principle |
UV-DOAS+PLS |
|||||||
Versions |
Regular range |
Low range |
Ultra-low range |
|||||
Ranges |
Min range |
Max range |
Min range |
Max range |
Min range |
Max range |
||
Component |
SO2 |
0-300ppm |
0-3000ppm |
0-100ppm |
0-300ppm |
0-50ppm |
0-100ppm |
|
NO |
0-300ppm |
0-3000ppm |
0-100ppm |
0-300ppm |
0-50ppm |
0-100ppm |
||
NO2 |
0-500ppm |
0-4000pp |
0-300ppm |
0-1000ppm |
0-100ppm |
0-200ppm |
||
O2 (ECD or ZrO2) |
0-5% |
0-25% |
0-5% |
0-25% |
0-5% |
0-25% |
||
Linearity |
≤±3%F.S. |
|||||||
Repeatability |
≤2% |
|||||||
Zero Drift |
≤±2%F.S./7d |
|||||||
Span Drift |
≤±2%F.S./7d |
|||||||
Response Time(T90) |
≤50s |
|||||||
Working Temperature |
-20℃~+50℃ |
|||||||
Preheating Time |
No preheating |
20min |
40min |
|||||
Sample Gas Interface |
Φ6 Bi-Lok |
|||||||
Sample Gas Flow |
Range: (0.5~2)L/min, fluctuation<25% |
|||||||
Sample Gas Pressure |
Current ambient pressure ±0.1Bar |
|||||||
Sample Gas Temp. |
0℃~+50℃ |
|||||||
Sample Gas Humidity |
No condensation (under sample gas temperature) |
|||||||
4-20mA Input |
3 , configurable, 100Ω load |
|||||||
4-20mA Output |
5, output content can be configured, maximum load capacity ≤800Ω |
|||||||
Switching value Input |
6 , configurable |
|||||||
Relay Output |
14, output content can be configured. DC30V2A |
|||||||
Digital I/O |
1×Rs232, 1×Rs485 (Support Modbus protocol) |
|||||||
Installation Method |
19' panel |
|||||||
Supply Voltage |
220VAC±10% |
|||||||
Rated Power |
About 100W |
About 120W |
Application Of EM-5 Flue Gas Analyzer
Coal-fired power plant
Cement plant
Industrial furnace
DeSOX process monitoring
DeNOX process monitoring
Waste incineration plant
PVC process in chlor-alkali plant
Titanium dioxide production process
Sulfur recovery process
Natural gas purification process
Methyl iodide analysis for the online gas analysis in coal chemical industry
On-line air monitoring
With features of high measurement accuracy, high reliability, fast response time, wide application scope, it can be widely applied in occasions of environment online monitoring, industrial process control, safety monitoring and etc.
Depended on years of laser gas analyzer and flue gas analyzer research experience and combined with practical experience of thousands of successful application cases, analyzers provided by continuous emission monitoring system suppliers like ZETIAN have derived multiple types (standard type, low emission type, ultra-low emission type, and customized type) to meet different index requirements in different working condition.
UV DOAS technology. The optical technology platform consists of a light source, gas chamber, optical fiber and spectrograph (including the diaphragm, a holographic grating, linear array detector) and other optical components.
Ultraviolet light emitted by light source enters the gas chamber by an optical window and absorbed by measured sample gas flowed through the gas chamber. The light carrying sample absorption information will be gathered by the lens and coupled into an optical fiber and then transmits into spectrometer through fiber. After light splitting and photo-voltaic conversion, the gas absorption spectrum is obtained.
The corresponding component concentration in the gas can be calculated by analyzing the spectrum.
Note: the difference among standard analyzer, low emission analyzer, and ultra-low analyzer is the different optical path length of gas chamber; for standard online gas analyser, it adopts dual lens collimation system with 0.25m optical path; for low emission analyzer, it adopts multiple return structure with 1m optical path; for ultra-low emission analyzer, it uses white cell structure.
Zetian Group has high-quality continuous gas analyzer for sale, contact us now.
HANGZHOU ZETIAN TECHNOLOGY is a professional air pollution monitoring instruments manufacturers, we provide Air Quality Monitor, Water Quality Analyzers, Continuous Emission Monitoring & Process Control, Gas Analysis Modules,online flue gas analyzer, water quality testing and etc. Want to know about flue gas analyser working principle? Please contact us.
Bonne affaire : acheter au vendeur
Cette page concerne les importateurs et exportateurs de EM-5 Flue Gas Analyzer Rechercher dans la catégorie : Informatique Rechercher dans la catégorie : flue, analyzer |
Dimanche 15 décembre 2024
Prix : 12,90 €
Maillot de foot avec short officiel MANCHESTER UNITED Tous les modèles Clubs Étrangers disponible sur notre site internet. Vendu pour le professionnel 12.90€ / pcs par 36 pièces dans le carton mélange de tailles et clubs possible. Ce tarif est valable pour 36 pcs, non négociable...
Samedi 25 octobre 2014
Thermo Farbgel UV Gel Farbwechsel Farbwechsel UV Gel Thermo UV Farbgel 10ml Thermo Gel Thermogel Farbwechsel UV-Farbgel Purpur Nail Art Farbwechsel UV Gel 12ml Purpur Thermo UV Gel Hochwertiges und Profiles Thermo Farbgel in Purpur von 10ml. Farbänderung bei Thermogelen werden...
Lundi 20 juillet 2015
Maroc vente fournisseurs, tissus technique, tapis isolation acoustique, grossistes, producteurs, import-export Société italienne spécialisée en production de tissus techniques pour construction et travaux publics cherche importateur exclusif au Maroc pour tapis isolation acoustique...